Aqua TROLL 500 / 600 multiparameter sonde (Modbus PLC interface) Modbus register map

Serial device: in Modbus Connect, choose “Add to existing device” and pick the serial (RTU) device you created on your COM port.

Default configuration

Protocol
Modbus RTU over RS-485
Unit ID
1
Addressing
0-based
Byte order
Big endian
Word order
High word first
Baud rate
19200
Parity
Even
Data bits
8
Stop bits
1

To read this device over Modbus RTU in Modbus Connect:

  1. In Modbus Connect, add a serial (RTU) device on your COM port with the serial settings and unit ID shown above.
  2. Click Add to Modbus Connect (or open the downloaded JSON), choose “Add to existing device” and pick that serial device. Its registers, byte order and word order are added; the serial device keeps its own line settings.

The “Add device” tab creates a Modbus TCP device (IP address and port), so use it only when you reach this device through a Modbus RTU-to-TCP gateway set to the serial settings above.

Register map

AddressCountNameTypeUnitScaleAccess
54502
Temperature
Parameter ID 1, holding register number 5451 (address 5450), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). The block tail is listed below as rows 5452-5456 of this seed; the other parameters repeat the same five registers at their own address + 2. Default unit: °C (the unit follows the unit-ID register of the block).
Float32°C1R
54521
Data quality ID, temperature block
Offset +2 of the temperature parameter block (5450). Value 0 means there are no errors or warnings.
UInt16-1R
54531
Unit ID, temperature block
Offset +3 of the parameter block: unit ID of the returned value, marked writable (R/W) in the block table although its Type column prints 'float' for this single-register row (the unit IDs are integers). Appendix C lists the IDs (1 = °C, 2 = °F, 3 = K, 17 = PSI, ... 65 = µS/cm, 145 = pH, 162 = mV, 177 = % saturation, 194 = NTU, 257 = RFU).
UInt16-1RW
54541
Parameter ID, temperature block
Offset +4 of the parameter block: the parameter ID this block belongs to (1 for temperature).
UInt16-1R
54552
Off-line sentinel value, temperature block
Offset +5 of the parameter block (2 registers): the value returned on error or when the parameter is not available. Marked R/W in the block table, so the sentinel can be written. The default sentinel is 0.0.
Float32°C1RW
54572
Pressure
Parameter ID 2, holding register number 5458 (address 5457), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: psi (the unit follows the unit-ID register of the block).
Float32psi1R
54642
Depth
Parameter ID 3, holding register number 5465 (address 5464), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: ft (the unit follows the unit-ID register of the block).
Float32ft1R
54712
Level, depth to water
Parameter ID 4, holding register number 5472 (address 5471), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: ft (the unit follows the unit-ID register of the block).
Float32ft1R
54782
Level, surface elevation
Parameter ID 5, holding register number 5479 (address 5478), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: ft (the unit follows the unit-ID register of the block).
Float32ft1R
55062
Actual conductivity
Parameter ID 9, holding register number 5507 (address 5506), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µS/cm (the unit follows the unit-ID register of the block).
Float32µS/cm1R
55132
Specific conductivity
Parameter ID 10, holding register number 5514 (address 5513), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µS/cm (the unit follows the unit-ID register of the block).
Float32µS/cm1R
55202
Resistivity
Parameter ID 11, holding register number 5521 (address 5520), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: ohm·cm (the unit follows the unit-ID register of the block).
Float32ohm·cm1R
55272
Salinity
Parameter ID 12, holding register number 5528 (address 5527), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: PSU (the unit follows the unit-ID register of the block).
Float32PSU1R
55342
Total dissolved solids
Parameter ID 13, holding register number 5535 (address 5534), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: ppt (the unit follows the unit-ID register of the block).
Float32ppt1R
55412
Density of water
Parameter ID 14, holding register number 5542 (address 5541), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: g/cm³ (the unit follows the unit-ID register of the block).
Float32g/cm³1R
55552
Barometric pressure
Parameter ID 16, holding register number 5556 (address 5555), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mmHg (the unit follows the unit-ID register of the block).
Float32mmHg1R
55622
pH
Parameter ID 17, holding register number 5563 (address 5562), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: pH (the unit follows the unit-ID register of the block).
Float32pH1R
55692
pH mV
Parameter ID 18, holding register number 5570 (address 5569), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mV (the unit follows the unit-ID register of the block).
Float32mV1R
55762
ORP
Parameter ID 19, holding register number 5577 (address 5576), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mV (the unit follows the unit-ID register of the block).
Float32mV1R
55832
Dissolved oxygen concentration
Parameter ID 20, holding register number 5584 (address 5583), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
55902
Dissolved oxygen, % saturation
Parameter ID 21, holding register number 5591 (address 5590), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: % (the unit follows the unit-ID register of the block).
Float32%1R
56112
Chloride (Cl-)
Parameter ID 24, holding register number 5612 (address 5611), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
56182
Turbidity
Parameter ID 25, holding register number 5619 (address 5618), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: NTU (the unit follows the unit-ID register of the block).
Float32NTU1R
56532
Oxygen partial pressure
Parameter ID 30, holding register number 5654 (address 5653), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: torr (the unit follows the unit-ID register of the block).
Float32torr1R
56602
Total suspended solids
Parameter ID 31, holding register number 5661 (address 5660), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
56672
External voltage
Parameter ID 32, holding register number 5668 (address 5667), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: V (the unit follows the unit-ID register of the block).
Float32V1R
56742
Battery capacity (remaining)
Parameter ID 33, holding register number 5675 (address 5674), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: % (the unit follows the unit-ID register of the block).
Float32%1R
56812
Rhodamine WT concentration
Parameter ID 34, holding register number 5682 (address 5681), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µg/L (the unit follows the unit-ID register of the block).
Float32µg/L1R
56882
Rhodamine WT fluorescence intensity
Parameter ID 35, holding register number 5689 (address 5688), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: RFU (the unit follows the unit-ID register of the block).
Float32RFU1R
56952
Chloride (Cl-) mV
Parameter ID 36, holding register number 5696 (address 5695), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mV (the unit follows the unit-ID register of the block).
Float32mV1R
57022
Nitrate as nitrogen (NO3 as N) concentration
Parameter ID 37, holding register number 5703 (address 5702), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
57162
Ammonium as nitrogen (NH4 as N) concentration
Parameter ID 39, holding register number 5717 (address 5716), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
57232
Ammonium (NH4) mV
Parameter ID 40, holding register number 5724 (address 5723), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). The manual's default-units cell for this row prints mg/L although the parameter is an mV electrode reading; the row keeps mV and records the discrepancy. Default unit: mV (the unit follows the unit-ID register of the block).
Float32mV1R
57302
Ammonia as nitrogen (NH3 as N) concentration
Parameter ID 41, holding register number 5731 (address 5730), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
57372
Total ammonia as nitrogen (NH3 as N) concentration
Parameter ID 42, holding register number 5738 (address 5737), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mg/L (the unit follows the unit-ID register of the block).
Float32mg/L1R
57792
Eh
Parameter ID 48, holding register number 5780 (address 5779), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: mV (the unit follows the unit-ID register of the block). The manual's default-units cell prints '162 = mg/L' here, the same abbreviation slip as the Ammonium mV row: unit ID 162 is mV in Appendix C.
Float32mV1R
57862
Velocity
Parameter ID 49, holding register number 5787 (address 5786), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). The default-units cell of the manual prints µg/L for this row, which cannot be a velocity unit (the unit column of that page drifts); no unit is claimed here.
Float32-1R
57932
Chlorophyll-a concentration
Parameter ID 50, holding register number 5794 (address 5793), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µg/L (the unit follows the unit-ID register of the block).
Float32µg/L1R
58002
Chlorophyll-a fluorescence intensity
Parameter ID 51, holding register number 5801 (address 5800), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: RFU (the unit follows the unit-ID register of the block).
Float32RFU1R
58212
Blue green algae - phycocyanin concentration
Parameter ID 54, holding register number 5822 (address 5821), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µg/L (the unit follows the unit-ID register of the block).
Float32µg/L1R
58282
Blue green algae - phycocyanin fluorescence intensity
Parameter ID 55, holding register number 5829 (address 5828), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: RFU (the unit follows the unit-ID register of the block).
Float32RFU1R
58492
Blue green algae - phycoerythrin concentration
Parameter ID 58, holding register number 5850 (address 5849), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: µg/L (the unit follows the unit-ID register of the block).
Float32µg/L1R
58562
Blue green algae - phycoerythrin fluorescence intensity
Parameter ID 59, holding register number 5857 (address 5856), 32-bit IEEE-754 float, high word at the lower address. Every parameter starts a 7-register block: value at +0 (2 registers), data-quality ID at +2, unit ID at +3, parameter ID at +4, off-line sentinel value at +5 (2 registers). Default unit: RFU (the unit follows the unit-ID register of the block).
Float32RFU1R
69831
Parameter availability bitmap, bits for parameter IDs 1-16
Part 1 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69841
Parameter availability bitmap, bits for parameter IDs 17-32
Part 2 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69851
Parameter availability bitmap, bits for parameter IDs 33-48
Part 3 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69861
Parameter availability bitmap, bits for parameter IDs 49-64
Part 4 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69871
Parameter availability bitmap, bits for parameter IDs 65-80
Part 5 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69881
Parameter availability bitmap, bits for parameter IDs 81-96
Part 6 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69891
Parameter availability bitmap, bits for parameter IDs 97-112
Part 7 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69901
Parameter availability bitmap, bits for parameter IDs 113-128
Part 8 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69911
Parameter availability bitmap, bits for parameter IDs 129-144
Part 9 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69921
Parameter availability bitmap, bits for parameter IDs 145-160
Part 10 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69931
Parameter availability bitmap, bits for parameter IDs 161-176
Part 11 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69941
Parameter availability bitmap, bits for parameter IDs 177-192
Part 12 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69951
Parameter availability bitmap, bits for parameter IDs 193-208
Part 13 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R
69961
Parameter availability bitmap, bits for parameter IDs 209-219
Part 14 of the 14-register sensor-discovery block that starts at register number 6984 (address 6983). Reading the block makes the sonde scan its sensor ports and update its parameter map; a set bit means that parameter ID is available.
UInt16-1R

Read this device live with Modbus Connect, free for Windows, macOS and Linux

Related guides

Notes

Source: In-Situ 'Aqua TROLL 500 Operator's Manual', part number 0096400, vendor-hosted PDF at https://in-situ.com/pub/media/support/documents/Aqua_TROLL_500_Manual.pdf (SHA-256 a61485d06e4d496dbd6da50ae14ade1ee8fe5ea254f53b16761a98e5f93b663c), chapter 'Modbus PLC Interface' (pages 35-37) and Appendices A 'Sensor Discovery', B 'Parameter Numbers and Locations' and C 'Unit IDs'; the Aqua TROLL 600 manual carries the same chapter. Setup frame from 'Programming the PLC', with the instrument reset to factory defaults first: Modbus RTU, 1 start bit, 19200 baud, 8 data bits, even parity, 1 stop bit, and the device address set to 1 - the manual prints no factory slave address of its own. Addressing: Appendix B prints a 'Holding Register Number' and a 'Holding Register Address' one lower (5451 / 5450), and the chapter notes that some PLCs use the number and others the address, so the address column is stored as the wire address ('zero_based'). Word order: every measurement is a 32-bit IEEE-754 float with the high word at the lower address (BigEndian + HighFirst). Firmware 1.71 or later is required, and a measurement sequence should begin with the 14-register discovery block at register 6984 so the sonde rescans its ports; it is seeded as rows 6983-6996. An unavailable parameter returns the block's sentinel value instead of an exception. Parameter IDs are fixed, so these addresses hold for any sensor set and the discovery block reports which are populated. Default units apply until the unit-ID register is changed. Three rows whose default-units cell disagrees with the parameter are explained in their own text: Velocity (prints ug/L), Ammonium (NH4) mV (prints mg/L) and Eh (prints '162 = mg/L', where 162 is mV). Appendix B also prints register number 5894 for Chlorophyll-a, while its own address column (5793) and formula (ID 50 -> 5794) agree on 5794, so address 5793 is stored.